Model 200 Series
Programmable Option
Users Manual
Przeglądanie stron 0
1 2 3 4 5 6 ... 14 15

Podsumowanie treści

Strona 1 - Users Manual

Model 200 SeriesProgrammable OptionUsers Manual

Strona 2

7KDS Programmable PumpsRS232 Commands and ResponsesAll commands and responses in standard pumps remain the same, however, the programmode does have ad

Strona 3

8KDS Programmable PumpsRS232 Commands and Responses (Continued)mode prgmsets pump in program mode. Response :number n sets number of steps in program.

Strona 4

9KDS Programmable PumpsRS232 Commands and Responses (Continued)timeleft?queries time remaining in active stepresponse: xx:xx:xx where hr:min:secnumber

Strona 5

10KDS Programmable PumpsRS232 Commands and Responses (Continued)PROGRAM EXAMPLESyringe selected, 4.70 mm diameterRS232 programmingmode prgm Select Pro

Strona 6

11KDS Programmable PumpsAs viewed from rear of the pump.Pin 1,6 Controlled Output High/Low states set at start of each step (could be used for valve c

Strona 7

Limited WarrantyKD Scientific Inc. warrants to the first consumer purchaser, for a period of one year from thedate of purchase that this unit, when s

Strona 8

EU Directives WEEE and RoHSTo Our Valued Customers:We are committed to being a good corporate citizen. As part of that commitment,we strive to maintai

Strona 9

ContentsGeneral Information ...1 Features .

Strona 10 - KDS Programmable Pumps

1KDS Programmable PumpsGeneral InformationKEYPAD PROGRAMMABLE PUMPSThe "program mode " is available on all Series 200 models and permits mul

Strona 11

2KDS Programmable PumpsFeaturesPARAMETERS WHICH CAN BE PROGRAMMED IN EACH STEPTime durationInfuse or withdrawStart flowrateEnd flowrateTTL output sett

Strona 12

3KDS Programmable PumpsMenu OperationAfter selecting MODE and then PROGRAM the display will show STEP which leads into edit-ing the program.1) Number

Strona 13

4KDS Programmable PumpsMenu Operation (Continued)5) Rate (Continued)c) Options:1) if R1=R2=0 The pump is stopped, no flowrate.2) if R1<R2 Flowrate

Strona 14

5KDS Programmable PumpsMenu Operation (Continued)8) Loop (Continued)b) LOOP COUNTAfter setting the initial step number of the loop, the menu will prom

Strona 15

6KDS Programmable PumpsRunning The ProgramRun The Run key starts the program; the displayed time counts down and the direction arrow flashes.Hold/Cont