Microliter OEMPump ModuleUsers ManualModel #78-2900Publication # 5417-004 Rev. B
10KDS Microliter OEM Module10OperationGETTING STARTEDPlug one end of the power cable into P8 connection on board. Connect communicationcable from PC t
11KDS Microliter OEM Module11HYPER-TERMINAL: Terminal Emulation SoftwareREMOTE CONTROL VIA THE RS-232 INTERFACE USING HYPERTERMINAL** Normally include
12KDS Microliter OEM Module12HYPER-TERMINAL: Terminal Emulation Software (continued)You may want to save the setup information under a descriptive fil
13KDS Microliter OEM Module13Appendix A: SpecificationsAccuracy ±0.35%Reproducibility ±0.1%Number of Syringes OneSyringe Sizes 0.5µL (min), 1 mL (max)
14KDS Microliter OEM Module14The pump is set to the following default parameters on power-up and after anexternal enable command.DEFAULT SETTINGSSyrin
15KDS Microliter OEM Module15Unimetrics Series 4000 & 5000Size Diameter10 µl 0.46 mm25 0.72950 1.031100 1.460250 2.300500 3.2601000 4.610SGE Scien
16KDS Microliter OEM Module160.5 µ 0.103 0.001 39.840 0.001 0.664 0.001 0.040 0.000 0.0001 µ 0.150 0.002 84.480 0.001 1.408 0.001 0.085 0.000 0.0012 µ
17KDS Microliter OEM Module17Appendix D: Serial CommunicationCOMMANDS, QUERIES AND RESPONSESAfter each transmission to the pump terminating with a CR
18KDS Microliter OEM Module18Appendix D: Serial Communicationratew nnnnn u/u Set withdrawal ratevoli nnnnn uu Set infusion target volume nnnnn is 0 to
19KDS Microliter OEM Module***Note: Adjustable force settings0% 20%40%60%80%7 lbs.min.ForceDue to differences in mechanical drug forces, we do notreco
EU Directives WEEE and RoHSTo Our Valued Customers:We are committed to being a good corporate citizen. As part of that commitment,we strive to maintai
20KDS Microliter OEM ModuleAppendix E: Optional – Daisy-ChainingThe RS-232 Connection to the first board in the daisy-chain is made through the 9-pin
21KDS Microliter OEM ModulePACKING LIST1) Main Unit2) Universal input 100/250 VAC, 50/60 Hz, 18 watt power supply3) Line Cord4) Pump-to-power adapter
Declaration o f ConformityIn Accordance with ISO/IEC Guide 22 and EN 45014Manufacturer: KD Scientific Inc.84 October Hill Rd.Holliston, MA 01746-1388U
3KDS Microliter OEM ModuleTable of ContentsSUBJECT PAGE NO.General Information - Warranty and Repairs . . . . . . . . . . . . . . . . . . . . . . . .
4KDS Microliter OEM ModuleGeneral InformationSerial NumberAll inquires concerning our product should refer to the serial number of the unit. Serialnum
5KDS Microliter OEM Module5General Safety SummaryPlease read the following safety precautions to ensure proper use of your modular syringepump. To avo
General Safety Summary (continued)PLACE PRODUCT IN PROPER ENVIRONMENTEnvironmental Conditions:- Indoor use only- Temperature 5˚C to 40˚C (40˚F to 104˚
7KDS Microliter OEM Module7IntroductionTHEORY OF OPERATION:The µl OEM Pump Module is designed as a highly precise, single-syringe infu-sion/withdrawal
8KDS Microliter OEM Module8Introduction (continued)STALL DETECTIONThe motor is monitored by an optical encoder to confirm the programmed movement. If
9KDS Microliter OEM Module94132InstallationLOADING THE SYRINGE1. Release the syringe pusher (4) by pressing the bronze button (3) on the side of the p