Kd-scientific 910 OEM Pump Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Pompy Kd-scientific 910 OEM Pump. KD Scientific 910 OEM Pump User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 23
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
Milliliter OEM Pump Module
Users Manual
Model #78-2910
Publication # 5417-003 Rev. B
Przeglądanie stron 0
1 2 3 4 5 6 ... 22 23

Podsumowanie treści

Strona 1 - Users Manual

Milliliter OEM Pump ModuleUsers ManualModel #78-2910Publication # 5417-003 Rev. B

Strona 2

11KDS Milliliter OEM Module11OperationGETTING STARTEDPlug one end of the power cable into P8 connection on board. Connect communicationcable from PC t

Strona 3

12KDS Milliliter OEM ModuleProtecting Small, Fragile SyringesThe “mL” OEM Pump Module will hold micro liter size syringes down to 0.5µl size. Thesesma

Strona 4

13KDS Milliliter OEM Module13HYPER-TERMINAL: Terminal Emulation SoftwareREMOTE CONTROL VIA THE RS-232 INTERFACE USING HYPERTERMINAL** Normally include

Strona 5

14KDS Milliliter OEM Module14HYPER-TERMINAL: Terminal Emulation Software (continued)You may want to save the setup information under a descriptive fil

Strona 6

15KDS Milliliter OEM Module15Appendix A: SpecificationsAccuracy ±1.0%Reproducibility ±0.5%Number of Syringes OneSyringe Sizes 0.5µL (min), 50/60 mL B-

Strona 7

16KDS Milliliter OEM Module16The pump is set to the following default parameters on power-up and after anexternal enable command.DEFAULT SETTINGSSyrin

Strona 8

17KDS Milliliter OEM Module17Appendix B: Syringe Inside DiameterBecton DickinsonPlastic “Plasticpak”SizeDiameter1 cc 4.78 mm3 8.665 12.0610 14.5020 19

Strona 9

18KDS Milliliter OEM Module18nominal nominalsyringe diameter ul/hr ul/min ul/hr ml/minsize (mm) min max min max min max min max0.5 µ 0.103 0.001 39.84

Strona 10 - KDS Milliliter OEM Module

19KDS Milliliter OEM Module19Appendix D: Serial CommunicationCOMMANDS, QUERIES AND RESPONSESAfter each transmission to the pump terminating with a CR

Strona 11

20KDS Milliliter OEM Module20Appendix D: Serial Communication (continued)COMMANDS WITH NUMBERSratew nnnnn u/u Set withdrawal ratedia nn.nn Set syringe

Strona 12

EU Directives WEEE and RoHSTo Our Valued Customers:We are committed to being a good corporate citizen. As part of that commitment,we strive to maintai

Strona 13

21KDS Milliliter OEM Module***Note: Adjustable force settings0% 20%40%60%80%100%7 lbs.min.ForceDue to differences in mechanical drug forces, we do not

Strona 14

Appendix E: Optional – Daisy-ChainingThe RS-232 Connection to the first board in the daisy-chain is made through the 9-pin D-Subconnector or RJ-11 inp

Strona 15

PACKING LIST1) Main Unit2) Universal input 100/250 VAC, 50/60 Hz, 18 watt power supply3) Line Cord4) RS-232Communications cable5) I/O & E-Stop Cab

Strona 16

KDS Milliliter OEM ModuleDecla ration of Confo rmityIn Accordance with ISO/IEC Guide 22 and EN 45014Manufacturer: KD Scientific Inc.84 October Hill Rd

Strona 17

4KDS Milliliter OEM ModuleTable of ContentsSUBJECT PAGE NO.General Information - Warranty and Repairs . . . . . . . . . . . . . . . . . . . . . . . .

Strona 18

5KDS Milliliter OEM ModuleGeneral InformationSerial NumberAll inquires concerning our product should refer to the serial number of the unit. Serialnum

Strona 19

6KDS Milliliter OEM Module6General Safety SummaryPlease read the following safety precautions to ensure proper use of your modular syringepump. To avo

Strona 20

7KDS Milliliter OEM ModuleGeneral Safety Summary (continued)PLACE PRODUCT IN PROPER ENVIRONMENTEnvironmental Conditions:- Indoor use only- Temperature

Strona 21

8KDS Milliliter OEM Module8IntroductionTHEORY OF OPERATION:The “mL” OEM Pump Module is designed as a highly precise, single-syringe infusion/withdrawa

Strona 22

9KDS Milliliter OEM Module9Introduction (continued)STALL DETECTIONThe motor is monitored by an optical encoder to confirm the programmed movement. If

Strona 23 - Decla ration of Confo rmity

10KDS Milliliter OEM Module10InstallationLOCATION REQUIREMENTSFOR THE SYRINGE PUMP• A sturdy, level, clean and dry surface• Minimum of one inch (2.5cm

Komentarze do niniejszej Instrukcji

Brak uwag